Seite 2 - INTRODUCTION

?Power on but no display. Check the battery are in place and making good contact or correct polarity , replace a new battery or attach optional AC ad

Seite 3 - CONTROLS AND INDICATORS

The meter is warranted to be free from defects in material and workmanship for a period of one year from the date of purchase. This warranty covers no

Seite 4 - MODE OPTIONS

Congratulations on your purchase ofthe Hygro-Thermometer ! This uniqueinstrument is portable, battery operated in Temperature./Relative Humidity/Dew

Seite 5 - WET BULB FUNCTION

1. 2. "-" .Minus pressure display.3. MAX Maximum recorded.4. MIN Minimum recorded.5. HLD Hold current reading.6. td

Seite 6

Fig. A3Unit will turn itself off after 20 minutes(Sleep mode) To override Auto PowerOff function (Non-Sleep Mode) press ON/OFF and MIN/MAX button simu

Seite 7 - MAINTENANCE

Press and hold (td/WBT) button until "td" appears on left side of display. Unit will now display Dew Point Temp.See Fig.EPress "HOLD&q

Seite 8

(1) Press and hold MN/MX button until (MIN) appears on display.(See Fig.G) Display is now showing minimum humidity and temperature

Seite 9

6The meter is calibrated in house before shipping.To maintain the meter in the goodcondition for use , recommend tocalibrate the meter after long time

Seite 10 - TROUBLESHOOTING

If you want to calibration 75% , please remove the meter probe from the 33% salt bottle and plug it into the 75% salt bottle. Then press the "MN

Seite 11 - WARRANTY

Automatic Calibration :Turn off the meter , then plug the meterprobe into the 33% salt bottle and thenpower on the meter by pressing " ON/OFF bu